Troubleshooting

Problem

 

Description

Reaction

Display

 

 

 

MEM RST

Data Error

Calibrate device. Reconfigure any custom settings (alarm setpoints, data log, etc...)

PCB ERR #

Electronic component error

Contact MSA *)

PRG ERR #

Program memory error

Contact MSA *)

RAM ERR #

RAM error

Contact MSA *)

UNK ERR #

Unknown error

Contact MSA *)

SNS ERR #

Sensor error

Contact MSA *)

Sensor warning

Sensor is near the end of its life

&

Sensor alarm

Sensor has reached the end of its life and cannot be calibrated.

Replace sensor and recalibrate.

Battery warning (no alarm)

Remove from service as soon as possible and replace battery

BATT ERR

Battery empty (with alarm – flash LEDs, horn)

Device is no longer sensing gas.

Remove from service and replace battery.

Device does not turn ON

Low battery

Replace battery

*) If the error occurs during the warranty period, please contact the MSA customer service. Otherwise, the device must be put out of operation.

#: Some errors will show an additional error code number which should be noted and be available when contacting MSA.