Calibration

The device can be calibrated either manually using this procedure or automatically using the GALAXY Test Stand ( Calibrations).

Calibration must be performed using a flow regulator with a flow rate set to 0.25 liters per minute.

If a battery charging cycle is interrupted before it is completed (4 hours for a fully discharged battery), allow the device’s internal temperature to stabilize for 30 minutes before performing a Calibration.