Calibration Setup

The user can change and set the calibration values for each sensor.

It is also possible to

select whether the Cal Due screen is displayed
set the number of days until the next calibration is due
select whether the device password screen is displayed to protect Bump and Calibration operations

For more information, see the flow charts in Calibrations.

Figure 5 Calibration Setup

1. To bypass this setup, press the q or p button; otherwise, continue as follows.
2. Press the button to enter the submenu.

The calibration gas concentration of the first sensor is shown.

3. Press the q or p button to change the value.
4. Press the button to store the value.

Setup screen for the next sensor is displayed.

5. Repeat the procedure for all other sensors.

After setting the last sensor, the user is prompted to set CALDUE.

6. Press the q or p button to enable or disable CALDUE.
7. Confirm with the button.
8. If CALDUE is set on, press the q or p button to select the number of days
9. Confirm with the button.
10. Press the q or p button to enable or disable the calibration password (CAL PW) screen.
11. Confirm with the button.
12. After confirmation, continue to Alarm setup.