Fresh Air Setup and Zero Calibration
To skip the Zero procedure and move directly to the Calibration Span procedure, push the p button. If no button is pushed for 30 seconds, the device prompts user to perform a Span calibration before returning to the Normal Operation mode. |
1. | Press and hold the p button in Normal Operation mode for three seconds. |
2. | If calibration lockout option is selected, enter password. |
ZERO CAL? screen is then displayed.
ZERO screen displays. |
|
|
|||
|
Alternatively press the q button to execute a Fresh Air setup (FAS). See Turning ON and Fresh Air Setup for more details. |
After Zero calibration completes, the device momentarily displays “ZERO PASS” or “ZERO ERR” along with the flag of any sensor that failed. |
|
|
During zero calibration the O2 sensor is also span calibrated to 20.8% O2 fresh air, adjusting the calibration curve as needed. During span calibration, the O2 sensor's accuracy is checked against a known oxygen gas concentration without adjusting the calibration curve. |