Span Calibration
To skip the Span calibration procedure, push the p button.
NOTE: If the SPAN calibration of the combustible sensor is skipped after a successful ZERO calibration, the combustible sensor reading may be replaced with a moving display of “PLEASE WAIT” for a few moments. This is normal, and the device is fully operational once a combustible gas reading reappears.
If no button is pushed for 30 seconds, span calibration is skipped.
Because of the different possible combinations of gases that are possible, skipping a Span calibration could advance the user to the Span calibration of another installed sensor, or back to Measuring mode.
When calibrating with combustible gases > 100 % LEL, select the "Yes" option to prompt "Span Calibration?" BEFORE applying gas to the device.
|
1.
|
Connect one end of tubing to the cylinder regulator (supplied in the calibration kit). |
|
2.
|
Connect the other end of the tubing to the pump inlet. |
|
|
|
|
3.
|
Press the q button to calibrate (span) the device. |
|
○
|
“SPAN CALIBRATION“ flashes |
|
○
|
SPAN calibration starts. |
|
○
|
A progress bar shows the user how much of the calibration has already been completed. |
|
|
|
|
NOTE: After the SPAN calibration is completed, the device displays either
|
•
|
“SPAN CALIBRATION PASS” |
or
|
•
|
“SPAN CALIBRATION FAIL” |
NOTE: The device returns to Measuring mode.
|
|
|
| |
|
|
If a sensor is nearing its end-of-life, the "PASS" display is followed by the Sensor Life indicator © display.
|
•
|
While the sensor is still fully functional, this warning gives the user time to plan for a replacement sensor to minimize downtime. |
|
•
|
The © indicator blinks as the device returns to Measure mode. |
|
•
|
After 15 seconds, the blinking stops, but the © indicator continues to display during ongoing operations as a reminder of a sensor’s pending end-of-life. |
If a span calibration fails:
|
•
|
The Sensor Life Indicator © blinks to show a sensor has reached its end-of-life and should be replaced. |
|
•
|
The device remains in the Sensor Life alarm condition until the p button is pressed. |
|
•
|
After the alarm is cleared, the device enters Measure mode and the Sensor Life indicator © blinks during ongoing operations until the sensor is replaced and/or successfully calibrated. |
Span calibration can fail for reasons other than a sensor at the end of its life. If a span calibration failure occurs, verify items such as:
|
•
|
sufficient gas remaining in the calibration cylinder |
|
•
|
integrity of calibration tubing/fittings, etc. |
|
4.
|
Reattempt the span calibration before replacing the sensor. |