Calibration Setup
|
CALIBRATION OPTIONS
|
|
|
The Calibration Options menu has provisions to:
|
•
|
modify the calibration cylinder settings (CYLINDER SETUP) |
|
•
|
enable/disable calibration due and to set the number of days (CAL DUE OPTIONS) |
|
•
|
enable/disable the option to show the last cal date at turn on and (LAST CAL DATE) |
When enabled, the date of the last device calibration displays during the turn-ON process.
|
•
|
enable/disable the option for password protected calibration (CAL PASSWORD) |
When enabled, the device setup password must be entered prior to calibration.
Press:
|
•
|
the q button go to next page |
|
•
|
the p button to go previous page |
|
•
|
the button to enter setup. |
Setting Calibration Cylinder
This option has a dialog similar to the span calibration dialog.
The display shows all active sensors.
|
1.
|
Press the button to enter setup. |
The screen for the first calibration cylinder displays.
the q or p button to change the value.
the
button to confirm the setup.
With this confirmation the device automatically moves to the next cylinder setting.
|
3.
|
Repeat the sequence for changing the required settings for all necessary gas values. |
After the last setting is performed, the device returns to the Calibration Options menu.
|
|
The only allowed calibration gas for the 0-2000 ppm PID sensor is 100 ppm isobutylene balanced in air. Higher concentrations can cause false readings of the CO sensor.
|
Setting Cal Due Options
|
1.
|
Press the button to enter setup. |
|
2.
|
Press the q or p button to enable/disable this option. |
|
3.
|
Press the button to confirm. |
|
4.
|
After confirmation the device prompts the user to enter the number of days for the reminder. |
|
5.
|
Change number of days by pressing the q or p button. |
|
6.
|
Press the button to go to the next menu. |
Setting Last Cal Date
|
1.
|
Press the button to enable/disable this option. |
|
2.
|
Press the q button to go to the next page. |
|
3.
|
Press the p button to go to the previous page. |
Setting Calibration Password
|
1.
|
Press the button to enable/disable this option. |
|
2.
|
Press the q button to go to the next page. |
|
3.
|
Press the p button to go to the previous page. |
Back To Main Menu
|
1.
|
Press the button to go to Device Setup Menu |
The Cal Options screen displays
|
2.
|
Press the q button to go to the next (Alarm options) or the p button to exit the Setup menu. |