Factory-Set Alarm Thresholds and Setpoints

Check the device or calibration certificate for exact alarm levels as they vary depending on national or corporate regulations.

Sensor

LOW alarm

HIGH alarm

SETPOINT min

SETPOINT max

STEL

TWA

CL2

0.5 ppm

1.0 ppm

0.3 ppm

7.5 ppm

1.0 ppm

0.5 ppm

ClO2

0.1 ppm

0.3 ppm

0.1 ppm

0.9 ppm

0.3 ppm

0.1 ppm

CO

25 ppm

100 ppm

7 ppm

1700 ppm

100 ppm

25 ppm

CO-HC

25 ppm

100 ppm

10 ppm

8500 ppm

100 ppm

25 ppm

COMB

10 % LEL

20 % LEL

5 % LEL

60 % LEL

--1

--1

H2S

10 ppm

15 ppm

2 ppm

175 ppm

15 ppm

10 ppm

H2S-LC

5 ppm

10 ppm

0.3 ppm

70 ppm

10 ppm

1 ppm

HCN

4.5 ppm

10.0 ppm

2.0 ppm

20.0 ppm

10 ppm

4.5 ppm

NH3

25 ppm

50 ppm

10 ppm

75 ppm

35 ppm

25 ppm

NO

25 ppm

75 ppm

15 ppm

100 ppm

25 ppm

25 ppm

NO2 (S 20)

2.0 ppm

5.0 ppm

1.0 ppm

17.5 ppm

5.0 ppm

2.0 ppm

NO2 (XCell)

2.5 ppm

5.0 ppm

0.5 ppm

47.5 ppm

5.0 ppm

2.5 ppm

O2

19.5 %

23.0 %

5.0 %

24.0 %

--1

--1

PH3

0.3 ppm

1.0 ppm

0.3 ppm

3.75 ppm

1.0 ppm

0.3 ppm

PID

50 ppm

100 ppm

2 ppm

1500 ppm

25 ppm

10 ppm

SO2

2.0 ppm

5.0 ppm

0.5 ppm

17.5 ppm

5.0 ppm

2.0 ppm

1STEL and TWA not applicable for combustible and oxygen gases.

In environments with >100 %  LEL combustible gas present, devices with a catalytic combustible LEL sensor will be in a latching over-range alarm.