Troubleshooting
|
Error State |
Details |
Recommended Action |
|
Alternating display |
|
|
|
ADC ERROR |
Analogue measurement error |
Contact MSA |
|
MEM ERROR |
Memory error |
Contact MSA |
|
PROG ERROR |
Program error |
Contact MSA |
|
RAM ERROR |
RAM error |
Contact MSA |
|
BT ERROR |
Bluetooth error |
Contact MSA |
|
LOW BATTERY
(flashing) |
Battery Warning repeats every 30 seconds |
Remove from service as soon as possible and recharge or replace battery |
|
BATTERY ALARM |
Battery is completely discharged |
Device is no longer sensing gas; Remove from service and recharge or replace battery. |
|
Device does not turn ON |
Battery fully discharged |
Remove from service as soon as possible and -recharge or replace battery pack. |
|
SENSOR MISSING |
Sensor damaged or missing |
Replace sensor |
|
NO SENSORS |
No sensors are enabled |
Device must have at least one sensor enabled at all times |
|
|
Sensor warning |
Sensor is near the end of its life |
|
(flashing) |
Sensor alarm |
Sensor has reached the end of its life and cannot be calibrated. Replace sensor and recalibrate. |
|
PUMP ERROR |
Pump malfunction or flow path blockage |
Check flowpath for blockage. If error persists, remove from service. |
|
INVALID CONFIGURATION |
Sensor(s) installed in incorrect location. |
Install sensors as shown in Figure 18 . |