Zero Calibration Procedure
ZERO screen displays. To skip the ZERO procedure and move directly to the span calibration procedure, push the p button. If no button is pushed for 30 seconds, the device prompts user to perform a SPAN calibration before device returns to the Normal Measurement page. To perform ONLY a Fresh Air Setup at this time, press the |
|
|
|||
The message "SENSOR REFRESH" displays, followed by the message "ZERO CALIBRATION". The "REFRESH" message does not appear if a catalytic combustible sensor is not installed. ZERO calibration starts. A progress bar shows the user how much of the calibration has been completed. During the first moments of a ZERO calibration, the combustible sensor reading may be replaced by a moving display of “PLEASE WAIT”. This is normal. |
|
|
|||
|
After the ZERO calibration is completed the device displays either “ZERO CALIBRATION PASS” or “ZERO CALIBRATION FAIL”. |
|
|
|||
|
|
|||||
|
Only if the device passes the zero calibration the SPAN screen displays. |
|
|
button. The device then performs a Fresh Air Setup as described in