Specifications

Instrument Specifications

OPERATING TEMPERATURE RANGE

-20 to 50°C (-4 to +122°F)

Internal vibrator operates to 0°C (32°F)

HUMIDITY 10 to 95% RH, Non condensing
INGRESS PROTECTION IP67
RECOMMENDED STORAGE 0 to 40°C (32° to 104°F)
WARRANTY 2 years after activation, maximum or 18 hours alarm time when activated within 6 months after date of manufacture
AUDIBLE ALARM 95 dB typical
APPROXIMATE SIZE

3.4" H x 2.0" W x 1" D

(8.6 cm H x 5.1 cm W x 2.5 cm D)

WEIGHT 4 oz (113 grams)
SENSOR Electrochemical Sensors
SENSOR DETECTION RANGE H2S CO O2
  0-100 ppm 0-500 ppm 0-25% by volume
FACTORY-SET * ALARM SETPOINTS LOW ALARM HIGH ALARM
  CO 25 PPM 100 PPM
  H2S 10 PPM 15 PPM
  O2 19.5% 23.0%
BATTERY Lithium, non-rechargeable
* Other setpoints available upon request. They may also be modified prior to activation through the button or at any time via MSA FiveStar Link Software.

NOTE: This instrument has not been classified for use in atmospheres containing >21% oxygen.

Event Logging Specifications

NUMBER OF SHARED EVENTS 25 (most recent occurrences)
DATA TRANSMISSION METHOD Via MSA infrared adapter on a PC using MSA FiveStar Link Software > version 4.3
EVENT LOG INFORMATION
Alarm - Alarm Type - Alarm Value - Time/Date
 
Alarm Clear - Alarm Type - Alarm Value - Time/Date
 
Cal (Pass/Fail) - Time/Date
 
Bump (Pass/Fail) - Time/Date
 
Error Non-Shutdown - Error Type (See Error List) - Time/Date
 
End of Life - Reason (Error - see Error List) - Alarm minutes - Months Life - Time/Date.
TRANSMISSION TIME Typically < 60 seconds max.