Analog Communication

The device is capable of the following types of analog outputs:

4–20 mA, isolated, current sourcing
0–10 Vdc for gas sample point identification

Scale Analog Output: 4–20 mA

Full scale is 1000 ppm.

Normal gas readings are at 4–20 mA. The reading is proportional to the detected gas concentration.

For the factory default of 100% full scale setting (0.016 mA = 1 ppm):

Output (mA)

Gas Concentration (ppm)

4

0

5

63

6

125

7

188

8

250

9

313

10

375

11

438

12

500

13

563

14

625

15

688

16

750

17

813

18

875

19

937

20

1000

Analog Output Settings

Warm-up

Default = 3.5 mA

User adjustable between 0 and 3.75 mA

Calibration (nonconfigurable)

3.0 mA

Fault

Default = 2.0 mA

User adjustable between 0.5 and 3.75 mA

Device over range (nonconfigurable)

21 mA

0–10 Vdc

The voltage output can be used to identify any active gas sample line.

Sample Line

Vdc
4 or 8 pt system

Vdc
16 pt system

1

1

1.0

2

2

1.5

3

3

2.0

4

4

2.5

5

5

3.0

6

6

3.5

7

7

4.0

8

8

4.5

9

-

5.0

10

-

5.5

11

-

6.0

12

-

6.5

13

-

7.0

14

-

7.5

15

-

8.0

16

-

8.5

While the device is in calibration mode, the output is 10 Vdc.