Technical Specifications / Equipment Ratings
| 
                                                             Usage  | 
                                                        
                                                             Indoor use only  | 
                                                        |
| 
                                                             Overvoltage  | 
                                                        
                                                             Category II  | 
                                                        |
| 
                                                             Pollution  | 
                                                        
                                                             Degree 2  | 
                                                        |
| 
                                                             ac Power  | 
                                                        
                                                             100–240 Vac, 50/60 Hz, 200 VA maximum Supply voltage not to fluctuate more than ±10%  | 
                                                        |
| 
                                                             Operating Temperature  | 
                                                        
                                                             32–122°F (0 to 50°C)  | 
                                                        |
| 
                                                             Operating Altitude Limit  | 
                                                        
                                                             9762 ft (2975 m)  | 
                                                        |
| 
                                                             Operating Humidity  | 
                                                        
                                                             0 to 95%, noncondensing  | 
                                                        |
| 
                                                             Ingress Protection  | 
                                                        
                                                             IP54  | 
                                                        |
| 
                                                             Size  | 
                                                        
                                                             Height x Width x Depth = 15.6 in. x 13.6 in. x 7.6 in. (39.6 cm x 34.5 cm x 19.3 cm)  | 
                                                        |
| 
                                                             Weight  | 
                                                        
                                                             4 point unit = 14 lb (6.4 kg) 8 point unit = 15 lb (6.8 kg) 16 point = 17 lb (7.7 kg)  | 
                                                        |
| 
                                                             Sensor Type  | 
                                                        
                                                             Photoacoustic Infrared  | 
                                                        |
| 
                                                             Sample Line Tubing  | 
                                                        
                                                             ¼ in. OD x 1/8 in. ID  | 
                                                        
                                                             Polyurethane (ether based) or metal with tubing  | 
                                                    
| 
                                                             6 mm OD x 4 mm ID  | 
                                                    ||
| 
                                                             Sample Line Length  | 
                                                        
                                                             Up to 400 ft (121 m) for optimum performance, will support up to 1200 ft (366 m)  | 
                                                        |
| 
                                                             Gases Chillgard 5000  | 
                                                        
                                                             R-11, R-12, R-22, R-123, R-134A, R-401A, R-404A, R-407A, R-407C, R-407F, R-410A, R-422A, R-422D, R-427A, R-454B, R-466A, R-513A, R-514A, R-515B, R-1233zd(E), R-1234yf, R-1234ze, R-125, R-143A, R-152A, R-21, R-23, R-32, R-218, R-227, R-236FA, R-424A, R-426A, R-438A, R-448A, R-449A, R-452B, R-455B, R-507A, R-508B  | 
                                                        |
| 
                                                             Gases Chillgard 5000  | 
                                                        
                                                             NH3  | 
                                                        |
| 
                                                             Measuring Range  | 
                                                        
                                                             0–1000 ppm  | 
                                                        |
| 
                                                             Sensitivity (minimum detection)  | 
                                                        
                                                             1 ppm for all listed R-gases 10 ppm for NH3  | 
                                                        |
| 
                                                             Accuracy (linearity)  | 
                                                        
                                                             0– 50 ppm ±1 ppm, 51–1000 ppm ±10% of reading  | 
                                                        |
| 
                                                             Sensor Response Time  | 
                                                        
                                                             90% of gas reading < 19 sec 90% of gas reading < 60 sec for NH3  | 
                                                        |
| 
                                                             Relays  | 
                                                        
                                                             5 Form C, 8A 250 Vac resistive SPDT  | 
                                                        |
| 
                                                             Audible Alarm  | 
                                                        
                                                             95 ±5 dB(A) at 24 in. (61 cm) maximum  | 
                                                        |
| 
                                                             Communication Outputs  | 
                                                        
                                                             4–20 mA sourcing, 250 Ohm load; 0–10 V, 2 K Ohm minimum load; RS485 Modbus RTU;   | 
                                                        |
| Approvals | 
                                                             CAN/CSA C22.2 No. 61010-1-12 UL 61010-1, 3rd edition IEC/EN 61010-1:2010 CE LVD: 2014/30/EU EMC: 2014/35/EU RoHS: 2011/65/EU EN 14624:2012 R134A UKCA LVD: SI 2016 No. 1101 EMC: SI 2016 No. 1091 RoHS: SI 2016 No. 3032  |