Analog Output

Terminal Block

Connection Point

Block Name

Setting

P1

Term 7

0 - 20 mA

Analog Output

The 0 to 20 mA output is a current signal that corresponds to the following:

* HART (1.25 mA) is the default.

Analog Output

HART (1.25 mA)*

HART (3.5 mA)

HART Disabled/OFF

Startup

1.25 mA

3.5 mA

0 to 0.2 mA

FAULT Signal

1.25 mA

3.5 mA

0 to 0.2 mA

Test Mode

1.5 mA

3.5 mA

1.5 ± 0.2 mA

COPM Fault Signal

2.0 mA

3.5 mA

2.0 ± 0.2 mA

Ready Signal

4.05 ± 0.2 mA

4.05 ± 0.2 mA

4.05 ± 0.2 mA

INSTANT ALARM Signal

16.0 ± 0.2 mA

16.0 ± 0.2 mA

16.0 ± 0.2 mA

SUSTAINED ALARM Signal

20.0 ± 0.2 mA

20.0 ± 0.2 mA

20.0 ± 0.2 mA

Bluetooth Load 3.0 mA 3.0 mA 3.0 mA

NOTE: Refer to the FL5000 Modbus or HART manual for Current Range selection.

The maximum analog output load is 600 Ω.
The COPM Fault Signal may also be set to 0 mA at the factory (non HART only).
Startup mode lasts exactly 15 seconds.