Appendix: Specifications
Table 19 Specifications
Sensor Options |
||||||||||
CO |
H2S |
SO2 |
CL2 |
NO2 |
H2S 500 PPM |
Hydrogen 1000 PPM | NH3 - 100 PPM | NH3 - 1000 PPM | SO2 - 100 PPM | |
|---|---|---|---|---|---|---|---|---|---|---|
|
Operating Range2 |
-40 °C to +60 °C |
-40 °C to +60 °C |
-40°C to +60°C |
-40°C to +60°C |
20°C to +50°C* | -20 °C to +40 °C | -20°C to +50°C* | -30 °C to +50 °C | -30 °C to +50 °C | -20 °C to +50 °C |
|
Storage |
-40 °C to +60 °C |
-40 °C to +60 °C |
-40°C to +60°C |
-40°C to +60°C |
-40°C to +50°C | -40°C to +50°C | -0°C to +20°C | -20°C to +40°C | -20°C to +40°C | -20°C to +40°C |
|
Zero Drift1 |
< 1% FS per year |
< 1% FS per year |
≤ 1% FS per month |
< 1% FS per month |
< 1% FS per month | < 1% FS per month | < 1% FS per month | < 1% FS per month | < 1% FS per month | < 1% FS per month |
|
Span Drift1 |
< 2% FS per year |
< 2% FS per year |
≤ 2% FS per month |
≤ 2% FS per month |
≤ 2% FS per month | ≤ 2% FS per month | ≤ 2% FS per month | ≤ 2% FS per month | ≤ 2% FS per month | ≤ 2% FS per month |
|
Noise1 |
< 2 display units |
< 2 display units |
< 2 display units |
< 2 display units |
< 2 display units | < 2 display units | < 2 display units | < 2 display units | < 2 display units | < 2 display units |
|
Repeatability1 |
< ±2% |
< ±1% |
±1% |
±1% |
< ±10% | < ±10% | < ±10% | < 1% | < ±15% | < ±15% |
|
Resolution |
1 ppm |
0.1 ppm |
0.1 ppm |
0.1 ppm |
0.1 ppm | 1 ppm | 10 ppm | 1 ppm | 10 ppm | 0.5 ppm |
|
T90 |
< 9 s |
< 23 s |
< 6s |
< 12s |
< 60 s | < 60 s | <185 s | < 60 s | < 300 s | < 30 s |
| T50 | < 30 s | < 20 s | < 40 s | < 20 s | < 20 s | < 10 s | ||||
|
Humidity |
5-95% |
5-95% |
5-95% |
5-95% |
15-90% | 15-90% | 15-90% | 15-90% | 15-90% | 15-90% |
|
Expected Sensor Life |
5 years |
5 years |
5 years |
5 years |
2 years in the air | 2 years in the air | 2 years in the air | 5 years | 2 years in the air | 2 years in the air |
|
Warranty |
3 years |
3 years |
3 years |
3 years |
1 year | 1 year | 1 year | 3 years | 1 year | 1 year |
|
Enabling Technology |
XCell - non-consuming |
XCell - non-consuming |
XCell - non-consuming |
XCell - non-consuming |
Electro-chemical | Electro-chemical | Electro-chemical | Electro-chemical | Electro-chemical | Electro-chemical |
|
TruCal3 |
Yes |
Yes |
No |
No |
No |
No |
No |
No |
No |
No |
|
SafeSwap |
Yes |
Yes |
Yes |
Yes |
Yes | Yes | Yes | Yes | Yes | Yes |
|
Remote Mount Distance |
100 m |
100 m |
100 m |
100 m |
100 m | 100 m | 100 m | 100 m | 100 m | 100 m |
|
Power: Single Sensor |
2.8 W |
2.8 W |
2.8 W |
2.8 W |
2.8 W | 2.8 W | 2.8 W | 2.8 W | 2.8 W | 2.8 W |
|
Power: Dual Sensing |
3.6 W |
3.6 W |
3.6 W |
3.6 W |
3.6 W | 3.6 W | 3.6 W | 3.6 W | 3.6 W | 3.6 W |
| *Extended Range | None | -40°C to +50°C | None | -40°C to +60°C | None | -40°C to +60°C | ||||
| Sensor Options |
|||||
|
Oxygen |
Low Oxygen |
Combustible Catalytic |
XIR PLUS Combustibles |
XIR PLUS Carbon Dioxide |
|
| Operating Range2 |
-40 °C to +60 °C |
-20 °C to +50 °C |
-40°C to +60°C |
-40°C to +60°C |
-40°C to +60°C |
|
Storage |
-40 °C to +60 °C |
-20 °C to +40 °C |
-40°C to +60°C |
-40°C to +60°C |
-40°C to +60°C |
|
Zero Drift1 |
0.2% Vol per year |
< 1% FS per month |
< 5% LEL per year |
< 2% LEL per year |
< 2% FS per year |
|
Span Drift1 |
0.2% Vol per year |
< 2% FS per month |
< 5% LEL per year |
Not Applicable |
Not Applicable |
|
Noise1 |
< 2 display units |
< 2 display units |
1% LEL | 1% LEL |
0.05% |
|
Repeatability1 |
< ±1% |
< ±10% |
±1% |
±1% |
±1% |
|
Resolution |
01% Vol |
0.10% |
1% LEL |
1% LEL |
<0.05% |
|
T90 |
< 11 s |
< 30 s |
< 22 s |
< 2 s |
< 6 s |
|
Humidity |
10-95% |
10-90% |
5-95% |
5-95% |
5-95% |
|
Expected Sensor Life |
5 years |
2 years in air |
5 years |
10 years |
10 years |
|
Warranty |
3 years |
1 years |
3 years |
electronics: 5 years IR source bulb: 10 years |
|
|
Enabling Technology |
XCell - non-consuming |
Lead based |
XCell - GM cat bead |
XIR PLUS |
XIR PLUS |
|
TruCal3 |
No |
No |
No |
No |
No |
|
SafeSwap |
Yes |
Yes |
Yes |
No |
No |
| Remote Mount Distance | 100 m | 100 m | 100 m | 100 m | 100 m |
| Power: Single Sensor | 2.8 W | 2.8 W | 4.9 W | 6.7 W | 6.7 W |
| Power: Dual Sensing | 3.6 W | 3.6 W | 10.6 W | 11.6 W | 11.6 W |
| *Extended Range | -30 °C to +50 °C | ||||
Sensor Options |
||||||
HCL |
NO |
HCN |
ETO |
HF |
CL02 |
|
|---|---|---|---|---|---|---|
|
Operating Range2 |
-20 °C to +40 °C |
-30 °C to +40 °C |
-20 °C to +40 °C |
-20 °C to +40 °C |
0°C to +50°C* | -40 °C to +50 °C |
|
Storage |
-30 °C to +40 °C |
-20 °C to +40 °C |
-20 °C to +40 °C |
-20 °C to +40 °C |
0°C to +50°C* | -40°C to +60°C |
|
Zero Drift1 |
< 1% FS per month |
< 1% FS per month |
< 1% FS per month |
< 1% FS per month |
< 1% FS per month | < 1% FS per month |
|
Span Drift1 |
< 2% FS per month |
< 2% FS per month |
< 2% FS per month |
< 2% FS per month |
< 2% FS per month | < 2% FS per month |
|
Noise1 |
< 2 display units |
< 2 display units |
< 2 display units |
< 2 display units |
< 2 display units | < 2 display units |
|
Repeatability1 |
< ±35% |
< ±15% |
< ±15% |
< ±15% |
< ±15% | < ±15% |
|
Resolution |
1 ppm |
0.5 ppm |
1 ppm |
0.1 ppm |
0.1 ppm | 0.01 ppm |
|
T90 |
< 120 s |
< 20 s |
< 70 s |
< 140s |
< 90 s | < 30 s |
| T50 | < 30 s | < 5 s | < 30 s | < 50 s | < 60 s | <12s |
|
Humidity |
5 - 50%* |
15 - 90% | 15 - 90% | 15 - 90% | 15 - 60% | 10 - 95% |
|
Expected Sensor Life |
2 years |
2 years |
2 years |
2 years |
2 years | 5 years |
|
Warranty |
1 year | 1 year | 1 year | 1 year | 1 year | 3 years |
|
Enabling Technology |
Electrochemical |
Electrochemical | Electrochemical | Electrochemical | Electrochemical | XCell |
|
TruCal3 |
No |
No |
No |
No |
No |
No |
|
SafeSwap |
Yes |
Yes |
Yes |
Yes |
Yes | Yes |
|
Remote Mount Distance |
100 m |
100 m |
100 m |
100 m |
100 m | 100 m |
| Power Single Sensor | 2.8 W | 2.8 W | 2.8 W | 2.8 W | 2.8 W | 2.8 W |
| Power Dual Sensor | 3.6 W | 3.6 W | 3.6 W | 3.6 W | 3.6 W | 3.6 W |
| Extended Range | -30 C to 40 C | -30 C to 50 C | -30 C to 50 C | -10 C to 50 C | -40 C to 60 C | |
1 Typical response at room temperature
2 X5000 transmitter has an operating temperature range of -40°C to +60°C
3 Adaptive Environmental Compensation (AEC) and Diffusion Supervision (DS) will remain active between -20°C and +60°C.