Appendix: Specifications

Table 19 Specifications

Sensor Options

 

CO

H2S

SO2

CL2

NO2

H2S

500 PPM

Hydrogen 1000 PPM NH3 - 100 PPM NH3 - 1000 PPM SO2 - 100 PPM

Operating Range2

-40 °C to +60 °C

-40 °C to +60 °C

-40°C to +60°C

-40°C to +60°C

20°C to +50°C* -20 °C to +40 °C -20°C to +50°C* -30 °C to +50 °C -30 °C to +50 °C -20 °C to +50 °C

Storage

-40 °C to +60 °C

-40 °C to +60 °C

-40°C to +60°C

-40°C to +60°C

-40°C to +50°C -40°C to +50°C -0°C to +20°C -20°C to +40°C -20°C to +40°C -20°C to +40°C

Zero Drift1

< 1% FS per year

< 1% FS per year

≤ 1% FS per month

< 1% FS per month

< 1% FS per month < 1% FS per month < 1% FS per month < 1% FS per month < 1% FS per month < 1% FS per month

Span Drift1

< 2% FS per year

< 2% FS per year

≤ 2% FS per month

≤ 2% FS per month

≤ 2% FS per month ≤ 2% FS per month ≤ 2% FS per month ≤ 2% FS per month ≤ 2% FS per month ≤ 2% FS per month

Noise1

< 2 display units

< 2 display units

< 2 display units

< 2 display units

< 2 display units < 2 display units < 2 display units < 2 display units < 2 display units < 2 display units

Repeatability1

< ±2%

< ±1%

±1%

±1%

< ±10% < ±10% < ±10% < 1% < ±15% < ±15%

Resolution

1 ppm

0.1 ppm

0.1 ppm

0.1 ppm

0.1 ppm 1 ppm 10 ppm 1 ppm 10 ppm 0.5 ppm

T90

< 9 s

< 23 s

< 6s

< 12s

< 60 s < 60 s <185 s < 60 s < 300 s < 30 s
T50         < 30 s < 20 s < 40 s < 20 s < 20 s < 10 s

Humidity

5-95%

5-95%

5-95%

5-95%

15-90% 15-90% 15-90% 15-90% 15-90% 15-90%

Expected Sensor Life

5 years

5 years

5 years

5 years

2 years in the air 2 years in the air 2 years in the air 5 years 2 years in the air 2 years in the air

Warranty

3 years

3 years

3 years

3 years

1 year 1 year 1 year 3 years 1 year 1 year

Enabling Technology

XCell - non-consuming

XCell - non-consuming

XCell - non-consuming

XCell - non-consuming

Electro-chemical Electro-chemical Electro-chemical Electro-chemical Electro-chemical Electro-chemical

TruCal3

Yes

Yes

No

No

No

No

No

No

No

No

SafeSwap

Yes

Yes

Yes

Yes

Yes Yes Yes Yes Yes Yes

Remote Mount Distance

100 m

100 m

100 m

100 m

100 m 100 m 100 m 100 m 100 m 100 m

Power: Single Sensor

2.8 W

2.8 W

2.8 W

2.8 W

2.8 W 2.8 W 2.8 W 2.8 W 2.8 W 2.8 W

Power: Dual Sensing

3.6 W

3.6 W

3.6 W

3.6 W

3.6 W 3.6 W 3.6 W 3.6 W 3.6 W 3.6 W
*Extended Range       None -40°C to +50°C None -40°C to +60°C None   -40°C to +60°C
 

Sensor Options

 

Oxygen

Low Oxygen

Combustible Catalytic

XIR PLUS Combustibles

XIR PLUS Carbon Dioxide

Operating Range2

-40 °C to +60 °C

-20 °C to +50 °C

-40°C to +60°C

-40°C to +60°C

-40°C to +60°C

Storage

-40 °C to +60 °C

-20 °C to +40 °C

-40°C to +60°C

-40°C to +60°C

-40°C to +60°C

Zero Drift1

0.2% Vol per year

< 1% FS per month

< 5% LEL per year

< 2% LEL per year

< 2% FS per year

Span Drift1

0.2% Vol per year

< 2% FS per month

< 5% LEL per year

Not Applicable

Not Applicable

Noise1

< 2 display units

< 2 display units

1% LEL 1% LEL

0.05%

Repeatability1

< ±1%

< ±10%

±1%

±1%

±1%

Resolution

01% Vol

0.10%

1% LEL

1% LEL

<0.05%

T90

< 11 s

< 30 s

< 22 s

< 2 s

< 6 s

Humidity

10-95% 

10-90% 

5-95%

5-95%

5-95%

Expected Sensor Life

5 years

2 years in air

5 years

10 years

10 years

Warranty

3 years

1 years

3 years

electronics: 5 years

IR source bulb: 10 years

Enabling Technology

XCell - non-consuming

Lead based

XCell - GM cat bead

XIR PLUS

XIR PLUS

TruCal3

No

No

No

No

No

SafeSwap

Yes

Yes

Yes

No

No

Remote Mount Distance 100 m 100 m 100 m 100 m 100 m
Power: Single Sensor 2.8 W 2.8 W 4.9 W 6.7 W 6.7 W
Power: Dual Sensing 3.6 W 3.6 W 10.6 W 11.6 W 11.6 W
*Extended Range   -30 °C to +50 °C      

Sensor Options

 

HCL

NO

HCN

ETO

HF

CL02

Operating Range2

-20 °C to +40 °C

-30 °C to +40 °C

-20 °C to +40 °C

-20 °C to +40 °C

0°C to +50°C* -40 °C to +50 °C

Storage

-30 °C to +40 °C

-20 °C to +40 °C

-20 °C to +40 °C

-20 °C to +40 °C

0°C to +50°C* -40°C to +60°C

Zero Drift1

< 1% FS per month

< 1% FS per month

< 1% FS per month

< 1% FS per month

< 1% FS per month < 1% FS per month

Span Drift1

< 2% FS per month

< 2% FS per month

< 2% FS per month

< 2% FS per month

< 2% FS per month < 2% FS per month

Noise1

< 2 display units

< 2 display units

< 2 display units

< 2 display units

< 2 display units < 2 display units

Repeatability1

< ±35%

< ±15%

< ±15%

< ±15%

< ±15% < ±15%

Resolution

1 ppm

0.5 ppm

1 ppm

0.1 ppm

0.1 ppm 0.01 ppm

T90

< 120 s

< 20 s

< 70 s

< 140s

< 90 s < 30 s
T50 < 30 s < 5 s < 30 s < 50 s < 60 s <12s

Humidity

5 - 50%*

15 - 90% 15 - 90% 15 - 90% 15 - 60% 10 - 95%

Expected Sensor Life

2 years

2 years

2 years

2 years

2 years 5 years

Warranty

1 year 1 year 1 year 1 year 1 year 3 years

Enabling Technology

Electrochemical

Electrochemical Electrochemical Electrochemical Electrochemical XCell

TruCal3

No

No

No

No

No

No

SafeSwap

Yes

Yes

Yes

Yes

Yes Yes

Remote Mount Distance

100 m

100 m

100 m

100 m

100 m 100 m
Power Single Sensor 2.8 W 2.8 W 2.8 W 2.8 W 2.8 W 2.8 W
Power Dual Sensor 3.6 W 3.6 W 3.6 W 3.6 W 3.6 W 3.6 W
Extended Range -30 C to 40 C -30 C to 50 C -30 C to 50 C   -10 C to 50 C -40 C to 60 C

1 Typical response at room temperature

2 X5000 transmitter has an operating temperature range of -40°C to +60°C

3 Adaptive Environmental Compensation (AEC) and Diffusion Supervision (DS) will remain active between -20°C and +60°C.