Status Information
ULTIMA X5000 HART® Specification
Additional Device Status (Command #48)
The device specific status, which is returned via Common Practice Command #48, is shown in Table Additional Device Status . These bits may be set at power up to indicate an instrument failure or feature. They may also be set by a failure detected during continuous background diagnostic testing.
Table 2 Additional Device Status
| Byte | Bit | Name | Description |
|---|---|---|---|
| 0 | 0 | Configuration Fault | System configuration fault |
| 1 | Reserved | ||
| 2 | Reserved | ||
| 3 | Reserved | ||
| 4 | Reserved | ||
| 5 | Reserved | ||
| 6 | Reserved | ||
| 7 | General System Error | ||
| 1 | 0 | Low Supply Voltage Fault | The input power supply voltage is out of low end range. |
| 1 | RAM Checksum Fault | RAM test/check has failed. | |
| 2 | Flash Checksum Fault | Flash checksum check has failed. | |
| 3 | EEPROM Fault | EEPROM access or checksum check has failed. | |
| 4 | Internal Circut Fault | Internal circuitry fault or abnormal internal voltages (except battery) | |
| 5 | Sensor 1 Fault | Sensor 1 is in fault condition. | |
| 6 | Sensor 2 Fault | Sensor 2 is in fault condition. | |
| 7 | Relay Fault | One or more relays are in fault condition. | |
| 2 | 0 | LED Control Fault | LED control has failed. |
| 1 | Invalid EEPROM Event Log | Invalid EEPROM access event log index | |
| 2 | Invalid EEPROM Bluetooth | Invalid EEPROM access Bluetooth index | |
| 3 | Magnetic Switch Fault | Magnetic switch has been left closed for more than two minutes. | |
| 4 | Non-Critical Faults | Number of non-critical faults | |
| 5 | Reserved | ||
| 6 | Reserved | ||
| 7 | Reserved | ||
| 3 | 0 | EZ-Touch Switch Fault | EZ-Touch switch has been left closed for more than two minutes. |
| 1 | RTC Battery Low Fault | RTC backup battery low. | |
| 2 | Display/UI Board Fault | Communication or the power supply to display/UI board has failed. | |
| 3 | Fail to Complete Setup Fault | Base unit has been in setup mode for more than six minutes. | |
| 4 | Bluetooth Module Fault | Communication to Bluetooth module has failed. | |
| 5 | BCM Fault | Communication to BCM has failed. Or, BCM has reported failure. | |
| 6 | HART Reference Voltage Fault | HART® reference voltage has failed. | |
| 7 | HART Fault | Communication to HART® | |
| 4 | 0 | Device Specific Command 168 | Read Gas Name Device Specific Command (168) supported |
| 1 | Device Specific Command 202/203 | Fixed Current Mode Commands 202 & 203 supported | |
| 2 | Reserved | ||
| 3 | Reserved | ||
| 4 | Reserved | ||
| 5 | Reserved | ||
| 6 | Reserved | ||
| 7 | Reserved |