Status Information

ULTIMA X5000 HART® Specification

Additional Device Status (Command #48)

The device specific status, which is returned via Common Practice Command #48, is shown in Table Additional Device Status . These bits may be set at power up to indicate an instrument failure or feature. They may also be set by a failure detected during continuous background diagnostic testing.

Table 2 Additional Device Status

Byte Bit Name Description
0 0 Configuration Fault System configuration fault
1 Reserved
2 Reserved
3 Reserved
4 Reserved
5 Reserved
6 Reserved
7 General System Error
1 0 Low Supply Voltage Fault The input power supply voltage is out of low end range.
1 RAM Checksum Fault RAM test/check has failed.
2 Flash Checksum Fault Flash checksum check has failed.
3 EEPROM Fault EEPROM access or checksum check has failed.
4 Internal Circut Fault Internal circuitry fault or abnormal internal voltages (except battery)
5 Sensor 1 Fault Sensor 1 is in fault condition.
6 Sensor 2 Fault Sensor 2 is in fault condition.
7 Relay Fault One or more relays are in fault condition.
2 0 LED Control Fault LED control has failed.
1 Invalid EEPROM Event Log Invalid EEPROM access event log index
2 Invalid EEPROM Bluetooth Invalid EEPROM access Bluetooth index
3 Magnetic Switch Fault Magnetic switch has been left closed for more than two minutes.
4 Non-Critical Faults Number of non-critical faults
5 Reserved  
6 Reserved
7 Reserved
3 0 EZ-Touch Switch Fault EZ-Touch switch has been left closed for more than two minutes.
1 RTC Battery Low Fault RTC backup battery low.
2 Display/UI Board Fault Communication or the power supply to display/UI board has failed.
3 Fail to Complete Setup Fault Base unit has been in setup mode for more than six minutes.
4 Bluetooth Module Fault Communication to Bluetooth module has failed.
5 BCM Fault Communication to BCM has failed. Or, BCM has reported failure.
6 HART Reference Voltage Fault HART® reference voltage has failed.
7 HART Fault Communication to HART®
4 0 Device Specific Command 168 Read Gas Name Device Specific Command (168) supported
1 Device Specific Command 202/203 Fixed Current Mode Commands 202 & 203 supported
2 Reserved  
3 Reserved
4 Reserved
5 Reserved
6 Reserved
7 Reserved