Alarm Setup
|
ALARM OPTIONS |
|
|
The Alarm Options Menu allows the user to:
| • | enable/disable the vibrating alarm |
| • | enable/disable the audible alarm (horn) |
| • | enable/disable the Alarm LEDs |
| • | enable/disable the MOTIONALERT SELECTION page. |
If disabled, the user cannot change the device MotionAlert setting.
| • | set Sensor Alarms. |
Press
| • | the q button go to next page |
the p button to go previous page
the
button to enter setup.
Setting Vibrating Alarm
Press the
button to enable/disable this option.
Setting Horn Alarm
Press
button to enable/disable this option.
Setting LED Alarm
Press
button to enable/disable this option.
Setting MotionAlert Access
Setting this parameter allows the user to access the MOTIONALERT page from the MEASURE page.
If access is denied here:
| • | the user cannot access the MOTIONALERT page to enable or disable that feature |
| • | the InstantAlert feature ( Alarms) cannot be activated. |
| 1. | To grant or deny user access to the MOTIONALERT page, use the button to change the indicated selection. |
User access is:
permitted when the setting indicates ON.
denied when the setting indicates OFF.
| 2. | The selection is confirmed by pressing either the q or p button. |
Setting Sensor Alarms
This page allows modifying the preset alarm values of:
| • | LOW Alarm |
| • | HIGH Alarm |
| • | STEL Alarm |
| • | TWA Alarm. |
|
Factory set alarm levels are shown in Factory-Set Alarm Thresholds and Setpoints. |
| 1. | Press the button to enter Sensor Alarm setup. |
LOW Alarm Setup screen displays.
|
|
Figure 13 Sensor Alarm Setup
| 2. | Press |
the q button to abort the operation or
the p button to go to next alarm setup or
the
button to change the alarm setpoints.
Alarm Value for the first Sensor displays.
|
|
Figure 14 Sensor Alarm Setup
| 3. | Set values for Sensor Alarm by pressing the q or p button. |
| 4. | Press the button to confirm set value. |
| 5. | Repeat setting for all other sensors. |
| 6. | Press the p button to return to the Alarm Options menu. |
| 7. | Repeat setting for all other alarm types. |