Alarm Setup

ALARM OPTIONS

The Alarm Options Menu allows the user to:

enable/disable the vibrating alarm
enable/disable the audible alarm (horn)
enable/disable the Alarm LEDs
enable/disable the MOTIONALERT SELECTION page.

If disabled, the user cannot change the device MotionAlert setting.

set Sensor Alarms.

Press

the q button go to next page

the p button to go previous page

the button to enter setup.

Setting Vibrating Alarm

Press the button to enable/disable this option.

Setting Horn Alarm

Press button to enable/disable this option.

Setting LED Alarm

Press button to enable/disable this option.

Setting MotionAlert Access

Setting this parameter allows the user to access the MOTIONALERT page from the MEASURE page.

If access is denied here:

the user cannot access the MOTIONALERT page to enable or disable that feature
the InstantAlert feature ( Alarms) cannot be activated.
1. To grant or deny user access to the MOTIONALERT page, use the button to change the indicated selection.

User access is:

permitted when the setting indicates ON.

denied when the setting indicates OFF.

2. The selection is confirmed by pressing either the q or p button.

Setting Sensor Alarms

This page allows modifying the preset alarm values of:

LOW Alarm
HIGH Alarm
STEL Alarm
TWA Alarm.

Factory set alarm levels are shown in Factory-Set Alarm Thresholds and Setpoints.

1. Press the  button to enter Sensor Alarm setup.

LOW Alarm Setup screen displays.

Figure 13 Sensor Alarm Setup

2. Press

the q button to abort the operation or

the p button to go to next alarm setup or

the button to change the alarm setpoints.

Alarm Value for the first Sensor displays.

Figure 14 Sensor Alarm Setup

3. Set values for Sensor Alarm by pressing the q or p button.
4. Press the  button to confirm set value.
5. Repeat setting for all other sensors.
6. Press the p button to return to the Alarm Options menu.