Bump Test

A bump test quickly confirms that the gas sensors are functioning. Perform a full calibration to ensure accuracy.

There are two methods to perform a bump test on the ALTAIR io 4:

Automatically through the ALTAIR io DOCK
Manually using proper calibration gas, regulator, tubing, and a calibration cap.

Bump test frequency is often stipulated by national or corporate regulations; however, bump testing before each day's use is generally the accepted best safety practice. Perform a bump test more frequently if the device is subjected to physical shock or high levels of contaminants.

WARNING! 

Perform the bump test per the instructions in this user guide. Do not continue operation of a detector that is not able to pass a bump test.
Perform a bump test more frequently if the tested atmosphere contains the following materials, which may desensitize the combustible gas sensor and cause erroneous readings:
Organic silicones
Silicates
Lead-containing compounds
Hydrogen sulphide exposures over 200 ppm or exposures over 50 ppm for one minute.

Failure to follow these warnings can result in serious personal injury or death.

If the device fails the bump test, perform a calibration to check sensor accuracy per the Calibration section before using the device.

ALTAIR io DOCK Bump Test

To perform the automatic bump test:

1. Insert the device with the display facing the ALTAIR io DOCK Test Stand until it clicks.
2. The appropriate test runs automatically based on the Grid-configured settings and indicates bump test results upon completion.
3. Remove the ALTAIR io 4 by pressing the gray release button.

Manual Bump Test

The manual bump test requires the following equipment:

Proper, Unexpired Calibration Gas Cylinder
0.25 liters/min. Flow Regulator
1/8” ID Superthane Ester Tubing
Calibration Cap

To perform the manual bump test:

1. Verify the gas concentrations displayed match calibration gas cylinder.
2. Navigate to the bump test menu option.
3. Install the calibration cap.
a.   Insert the tab on the calibration cap into the left slot on the device.
b.   Press the calibration cap until it seats onto the device.
c.   Press both the top and right tab down onto the device until they snap into place.
d.   Ensure the calibration cap is properly seated.
e.   Connect one end of the tubing to the calibration cap.
f.   Connect the other end of the tubing to the gas regulator.
4. Select Start to begin the bump test.
5. Open the pressure reducer valve on the calibration gas cylinder.
6. The device will display the bump test results.
7. Close the valve after bump testing.
8. Remove the calibration cap.