Calibrating the ALTAIR Single Gas Detector
WARNING!
Perform a calibration per the instructions in this user guide. Do not continue operation of a detector that is not able to pass calibration.
Failure to follow this warning can result in serious personal injury or death.
|
•
|
While the ALTAIR Single Gas detector is designed to be a maintenance free instrument, under normal circumstances MSA recommends calibration at least every six months or in accordance with local regulations. The device must be calibrated if it does not pass a Bump Test or if any of the conditions listed below exist. |
|
•
|
For oxygen instruments, perform a calibration if any of the following occur: |
|
○
|
there are changes in barometric pressure (altitude changes) |
|
○
|
the instrument does not pass a bump test. |
|
•
|
For toxic instruments (CO and H2S), perform a calibration if any of the following occur: |
|
○
|
extended use in extreme temperatures |
|
○
|
high concentration exposure |
|
○
|
the instrument does not pass a bump test. |