Calibrating the ALTAIR Single Gas Detector

WARNING! 

Perform a calibration per the instructions in this user guide. Do not continue operation of a detector that is not able to pass calibration.

Failure to follow this warning can result in serious personal injury or death.

While the ALTAIR Single Gas detector is designed to be a maintenance free instrument, under normal circumstances MSA recommends calibration at least every six months or in accordance with local regulations. The device must be calibrated if it does not pass a Bump Test or if any of the conditions listed below exist.
For oxygen instruments, perform a calibration if any of the following occur: 
physical shock
there are changes in barometric pressure (altitude changes)
there are extreme changes in ambient temperature and humidity (see Instrument Specifications).
the instrument does not pass a bump test.
For toxic instruments (CO and H2S), perform a calibration if any of the following occur: 
physical shock
extended use in extreme temperatures
high concentration exposure
the instrument does not pass a bump test.